Model Number STRIP, GNP TMX 50CT MMMG/DL
Device Problem High Test Results (2457)
Patient Problems Muscle Weakness (1967); Blurred Vision (2137); Constipation (3274)
Event Date 09/08/2022
Event Type  Injury  
Event Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results from results obtained of 276, 238, 261, 233 and 222 mg/dl.The customer¿s expected blood glucose test results are 150 mg/dl fasting am, 198 mg/dl non-fasting afternoon and 189-199 mg/dl non-fasting evening.At the time of the call the customer reported feeling weak, constipated and having blurry vision.Medical attention was not needed at the time.During the call, a blood test was performed by the customer non-fasting and produced test result of 298 mg/dl using true metrix meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer¿s expiration date is 01/14/2023 and test strips were opened two weeks prior to call.The meter memory was reviewed for previous test result history: result 1: 276 mg/dl, date: 09-08, time: 08:59 am, non-fasting, result 2: 238 mg/dl, date: 09-07, time: 11:44 am, fasting, result 3: 261 mg/dl, date: 09-07, time: 11:43 am, fasting, result 4: 233 mg/dl, date: 09-07, time: 08:04 am, fasting, result 5: 222 mg/dl , date: 09-07, time: 02:00 am, fasting.
 
Manufacturer Narrative
Internal report reference number: (b)(4).Outcomes attributed to adverse event: adverse event report is being submitted due to symptoms related to diabetes: weak, blurry vision and constipation.Meter and test strips were not returned for evaluation.Note: manufacturer contacted customer in several follow-up calls to ensure the customer's condition had improved and that the replacement products resolved the initial concern - unable to establish contact with customer at this time.
 
Manufacturer Narrative
Sections with additional information as of 19-oct-2022: h6: updated fda¿s type, findings and conclusions codes.H10: meter was not returned for evaluation.Test strips were not returned for evaluation.Retention testing was performed using test strips from the same lot.Retention strip lot tested within specifications.Most likely underlying root cause: mlc-058: user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUE METRIX
Type of Device
SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key15495536
MDR Text Key300700084
Report Number1000113657-2022-00520
Device Sequence Number1
Product Code NBW
UDI-Device Identifier0002129200783
UDI-Public(01)0002129200783
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140100
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/14/2023
Device Model NumberSTRIP, GNP TMX 50CT MMMG/DL
Device Catalogue NumberRE4H01-81
Device Lot NumberZY4427S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date09/08/2022
Date Manufacturer Received09/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-