• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT CONFIRM; IMPLANTABLE CARDIAC MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT CONFIRM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM3500
Device Problems Signal Artifact/Noise (1036); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/09/2022
Event Type  malfunction  
Event Description
It was reported that following an episode of syncope, the event was not recorded in the device.Abbott technical support was contacted, the session records were analyzed, and noise was noted on the device.The episode was suspected to not have been recorded due to noise on the device.The physician elected to upgrade the device due to the condition of the patient and the device was explanted and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported events of ecg/iegms missing and signal artifact/noise could not be confirmed.The device was received in normal working conditions with battery voltage above elective replacement indicator (eri).Latest session record was analyzed which indicated no anomaly.Analysis performed, including sensing test at field and high sensitivity settings indicated normal device functionality.Longevity assessment was performed and device was in the normal range of operation with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15495885
MDR Text Key306082429
Report Number2017865-2022-39553
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/18/2022
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberP000123415
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-