SMITH & NEPHEW, INC. DBL OFFSET BROACH HDLE LT; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Model Number 71360089 |
Device Problem
Premature Separation (4045)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/07/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that, during a tha surgery a dbl offset broach hdle lt is loose.It is unknown how the procedure was completed.However; there was a delay greater than 30 min.The condition of the patient is unknown.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
H10 h6: the associated device was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure.The device does show signs of significant wear and use.A functional evaluation was conducted with a mating device and does confirms the stated failure.The handle is loose and worn, rendering the device inoperable.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|