ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2272
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fever (1858); Unspecified Infection (1930)
Event Date 09/12/2022
Event Type  Injury  
Manufacturer Narrative
A device history record (dhr) review was performed and review of the sterilization records confirmed normal sterilization cycles for the products.The device met specifications prior to leaving abbott manufacturing facilities.The product was returned and visual inspection was normal.The cause of infection could not be traced to the device. .
 
Event Description
Related manufacturer report number: 2017865-2022-40053, related manufacturer report number: 2017865-2022-40054.It was reported that the patient presented to the hospital with febrile illness.A transesophageal echocardiogram revealed vegetation of both the right ventricular and atrial leads.The patient was scheduled for system extraction, and the pulse generator, right atrial, and right ventricular leads were explanted.The patient was stable.
 
Manufacturer Narrative
A device history record (dhr) review was performed and review of the sterilization records confirmed normal sterilization cycles for the products.The device met specifications prior to leaving abbott manufacturing facilities.The product was returned and visual inspection was normal.The cause of infection could not be traced to the device. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15521355
MDR Text Key300987968
Report Number2017865-2022-40043
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734509589
UDI-Public05414734509589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Model NumberPM2272
Device Catalogue NumberPM2272
Device Lot NumberA000071184
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age69 YR
Patient SexMale
Patient Weight77 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-