Model Number URF-V
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/31/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found detachment of the insertion tube, watertightness of the insertion tube was not maintained, the bending angle was insufficient, forceps could not be inserted smoothly into the forceps channel, liquid leakage was found in the light guide bundle, the flexible tube of the insertion section was crushed, corrosion due to water leakage was found in the operation part, the neutral position was out of standard due to axial misalignment of the operating part, liquid leakage was observed in the grip, liquid leakage was found on the up/down plate, paint floating was found on the up/down plate, liquid leakage was found in the universal cord, and the universal cord was collapsed.Scratches were found on the grip, the universal cord, the rotating mechanism, the angle lever, the video cable, the video connector, the light guide connector, and the video connector case.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus, the uretero-reno videoscope had the coating torn and the metal coil exposed.Inspection and testing of the returned device found that the corrugated tube was falling off.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information based on the information supplied by the legal manufacturer.Please refer to the following sections for the new information: event date, relevant tests/laboratory data, including dates, concomitant products.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 11 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely due to stress, handling or other issues.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15521464
MDR Text Key306176032
Report Number3002808148-2022-02669
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170340321
UDI-Public04953170340321
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K072957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2022
Was the Report Sent to FDA? No
Date Manufacturer Received10/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/26/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CLV-S190, SERIAL NO.(B)(4).; OTV-S190, SERIAL NO.(B)(4).
-
-