Model Number D138502 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that an unknown patient underwent an unknown procedure with an 8.5f sheath with curve viz mdc.Doctor indicated that the valve is damaged.No patient consequences.To be more specific on what section broke/separated it was the hemostatic valve.The hemostasis valve (gasket) was not dislodged inside or outside the hub.The brim cap/hub did not become detached from the sheath.The sheath was not being used on the patient.Hemostatic valve separation is mdr-reportable.
|
|
Manufacturer Narrative
|
Initial reporter phone: (b)(6).This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|
Manufacturer Narrative
|
On 7-oct-2022, the bwi product analysis lab received the device for evaluation.The product investigation was subsequently completed.It was reported that an unknown patient underwent an unknown procedure with an 8.5f sheath with curve viz mdc.Doctor indicated that the valve is damaged.No patient consequences.Device evaluation details: visual analysis revealed that the hemostatic valve was dislodged inside the hub component.Microscopic examination of the hemostatic valve surface showed stress marks on the outer diameter.The stress marks suggest that excessive force or manipulation was applied due to an extreme off axis angle of insertion.Valve dislodgement occurs when extreme off axis angles are performed during insertion with the dilator, outside of what is recommended in the odp (optimal device performance guide) a device history record evaluation was performed for finished device number 50000106, and no internal actions related to the complaint were found during the review.The issue reported by the customer was confirmed.The odp contains the following caution: always insert a dilator straight into the center of the sheath¿s valve to prevent damage to the valve.Do not insert a dilator at an angle, as damage to the sheath valve may occur.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|
Search Alerts/Recalls
|
|
|