BIOSENSE WEBSTER INC 8.5F SHEATH WITH CURVE VIZ MDC; INTRODUCER, CATHETER Back to Search Results
Model Number D138502
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2022
Event Type  malfunction  
Event Description
It was reported that an unknown patient underwent an unknown procedure with an 8.5f sheath with curve viz mdc.Doctor indicated that the valve is damaged.No patient consequences.To be more specific on what section broke/separated it was the hemostatic valve.The hemostasis valve (gasket) was not dislodged inside or outside the hub.The brim cap/hub did not become detached from the sheath.The sheath was not being used on the patient.Hemostatic valve separation is mdr-reportable.
 
Manufacturer Narrative
Initial reporter phone: (b)(6).This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
On 7-oct-2022, the bwi product analysis lab received the device for evaluation.The product investigation was subsequently completed.It was reported that an unknown patient underwent an unknown procedure with an 8.5f sheath with curve viz mdc.Doctor indicated that the valve is damaged.No patient consequences.Device evaluation details: visual analysis revealed that the hemostatic valve was dislodged inside the hub component.Microscopic examination of the hemostatic valve surface showed stress marks on the outer diameter.The stress marks suggest that excessive force or manipulation was applied due to an extreme off axis angle of insertion.Valve dislodgement occurs when extreme off axis angles are performed during insertion with the dilator, outside of what is recommended in the odp (optimal device performance guide) a device history record evaluation was performed for finished device number 50000106, and no internal actions related to the complaint were found during the review.The issue reported by the customer was confirmed.The odp contains the following caution: always insert a dilator straight into the center of the sheath¿s valve to prevent damage to the valve.Do not insert a dilator at an angle, as damage to the sheath valve may occur.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8.5F SHEATH WITH CURVE VIZ MDC
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (IRVINE)
33 technology drive
irvine CA 92618
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-798
MDR Report Key15526286
MDR Text Key306508516
Report Number2029046-2022-02363
Device Sequence Number1
Product Code DYB
UDI-Device Identifier10846835016277
UDI-Public10846835016277
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K170997
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/08/2023
Device Model NumberD138502
Device Catalogue NumberD138502
Device Lot Number50000106
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received10/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-