MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-40 |
Device Problems
Migration or Expulsion of Device (1395); Infusion or Flow Problem (2964)
|
Patient Problems
Muscular Rigidity (1968); Cramp(s) /Muscle Spasm(s) (4521)
|
Event Date 09/22/2022 |
Event Type
Injury
|
Event Description
|
Information was received from a healthcare provider (hcp) via a company representative regarding a patient receiving lioresal (2000 mcg/ml at 1430 mcg/day) via an implanted pump.It was reported that the patient¿s symptoms returned on sunday.The patient¿s spasticity and spasms returned when lying flat on his back.The patient came to the er (emergency room) today and they tried a cap (catheter access port) study but were unable to get csf (cerebrospinal fluid).The patient was taken to surgery to have the catheter replaced.It was noted that the catheter tip was originally placed at the t7 level, however, today the tip was out of the spine.The anchor was still attached and the catheter length from anchor to tip was much shorter than what was expected.It was unknown if there were any environmental, external, or patient factors that may have led or contributed to the issue.The issue was noted to be resolved.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id: 8780, serial#: (b)(4), implanted: (b)(6) 2022, explanted: (b)(6) 2022, product type: catheter.Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 22-dec-2023, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: analysis of the catheter identified a kink in the catheter body.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|