Catalog Number EJ-05400-NRO |
Device Problem
Disconnection (1171)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/28/2022 |
Event Type
malfunction
|
Event Description
|
T was reported that 2 days after placement of the catheter, a leak was found.On checking, a nurse found that the catheter was completely disconnected from the snaplock adapter.Therefore, the catheter and the snaplock adapter were replaced with a new kit.According to the md, it was expected that some tension was applied due to the weight of the connected pca pump, but he/she confirmed that the snaplock adapter was locked to the end.
|
|
Manufacturer Narrative
|
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
|
Manufacturer Narrative
|
Qn#(b)(4).According to the engineer investigating the reported device failure, the customer reported that the snaplock assembly disconnected from the catheter.The customer returned one snaplock assembly nrfit and epidural catheter.The returned components were visually examined with and without magnification.Visual examination of the returned snaplock assembly revealed the snaplock appears typical with no defects or anomalies observed.Visual examination of the returned epidural catheter revealed the catheter appears used.Biological material can be seen between the catheter coils and adhesive residue is present on the catheter body exterior.No other defects or anomalies were observed.A functional leak test was performed on the returned snaplock assembly and epidural catheter.The returned epidural catheter was inserted from the proximal end into the returned snaplock assembly until it bottomed out and the snaplock assembly was closed.The components were confirmed to be secured by tugging gently.The snaplock assembly was connected to the lab leak tester and the pressure was increased to 10 psi to establish flow.The distal end of the catheter was capped off and the pressure was increased to 25 psi for 30 seconds.No leaks were detected, and the components remained secured.A functional spontaneous partial opening (spo) test was then performed.The proximal end of the epidural catheter was re-inserted into the snaplock assembly until it bottomed out and the snaplock assembly was locked.The components were confirmed to be secured by tugging gently on the catheter.The components were then left to sit for 72 hours in the locked position.After 72 hours, the snaplock assembly was confirmed to have remained securely locked with the catheter inserted.No functional issues were found.In summary, the reported complaint of the snaplock assembly disconnecting from the catheter could not be confirmed through functional testing of the returned snaplock assembly.The snaplock assembly was secured to the returned epidural catheter and passed the functional tests performed including a spontaneous partial opening (spo) test.A device history record review was performed on the epidural catheter and snaplock assembly with no relevant findings.The investigation found no evidence to suggest a manufacturing related cause.Therefore, based on functional testing of the returned sample, no problem was found with the returned sample.No further action is required at this time.
|
|
Event Description
|
It was reported that 2 days after placement of the catheter, a leak was found.On checking, a nurse found that the catheter was completely disconnected from the snaplock adapter.Therefore, the catheter and the snaplock adapter were replaced with a new kit.According to the md, it was expected that some tension was applied due to the weight of the connected pca pump, but he/she confirmed that the snaplock adapter was locked to the end.
|
|
Search Alerts/Recalls
|
|