Model Number 2217-50-006 |
Device Problems
Crack (1135); Device-Device Incompatibility (2919); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/23/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the impactors were worn and cracking.No surgical delay was reported.
|
|
Manufacturer Narrative
|
(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Additional information received indicated that the instrument was showing signs of crack on the colored impaction part/ unthreading from silver to attach to handle.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary: the device associated with this report was received for examination.Visual inspection of the returned device found the device overall appearance is worn.The reported worn condition was confirmed.However, no signs of cracks were found.A functional test performed with mating sample was unable to replicate the unable to assemble condition.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a date code was provided, which indicates that the device was manufactured on (9/2002).A manufacturing records evaluation (mre) was not performed since a valid finished good lot number was not provided for this device.Corrected: h3.
|
|
Search Alerts/Recalls
|