Model Number DM3500
Device Problems Migration or Expulsion of Device (1395); Under-Sensing (1661)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/16/2022
Event Type  malfunction  
Event Description
It was reported that the patient presented for follow-up in clinic.Upon interrogation, it was noted that the insertable cardiac monitor (icm) exhibited undersensing.It was noted that the icm migrated within the pocket and was causing the undersensing.Programming changes were performed.The patient was stable before, during, and after.
 
Event Description
Additional information received notes that the implantable cardiac monitor (icm) continued to exhibit false pause episodes.No changes were made.
 
Event Description
Additional information received notes that the implantable cardiac monitor (icm) continued to exhibit undersensing and have false pause episodes.No changes were made.
 
Event Description
Additional information received notes that programming changes were performed to resolve the issue.
 
Event Description
Additional information received notes that the implantable cardiac monitor (icm) had recurring issues of false pause episodes.No changes or interventions was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15572621
MDR Text Key306199808
Report Number2017865-2022-40351
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 05/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/06/2023
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberP000133313
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-