MAUDE Adverse Event Report: BIOSENSE WEBSTER INC CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ SMALL; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ SMALL; INTRODUCER, CATHETER Back to Search Results
Model Number D138501
Device Problems Material Separation (1562); Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2022
Event Type  malfunction  
Manufacturer Narrative
The product has not returned for analysis, however, pictures were provided by the customer to aid in the investigation.Evaluation is still in progress.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturers ref.#: (b)(4).
 
Event Description
It was reported that a patient underwent a cardiac ablation procedure with a carto vizigo¿ 8.5f bi-directional guiding sheath ¿ small and a hemostatic valve separation occurred.During preparation of carto vizigo¿ 8.5f bi-directional guiding sheath ¿ small, the nurse complained that she was not able to advance dilator into sheath.Physician tried as well but it was hard to advance dilator.During visible inspection it became obvious that sheath valve was dislocated into sheath handle and blocked inner lumen.The issue was noticed during preparation of the sheath.A new sheath was used successfully to treat patient without any other issues.The surgery was delayed 3 minutes due to the reported event.There were no patient consequences.
 
Manufacturer Narrative
If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4) on 8-dec-2022, the product investigation was reopened to process a correction as it was noticed that the following statement was incorrectly included for this product evaluation and the statement was removed from the product evaluation detailed conclusion: ¿however, due to the conditions observed in the hemostatic valve, an internal corrective action has been opened to address this issue.¿ this statement was previously reported in section h10 of the 3500a supplemental (follow-up) medwatch report # 2.Please consider this removed.
 
Manufacturer Narrative
On 18-oct-2022, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.Additionally, the manufactured date and expiration date have been provided.Therefore, fields d4.Expiration date and h4.Device manufacture date have been populated.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Manufacturer Narrative
It was reported that a patient underwent a cardiac ablation procedure with a carto vizigo¿ 8.5f bi-directional guiding sheath ¿ small and a hemostatic valve separation occurred.During preparation of carto vizigo¿ 8.5f bi-directional guiding sheath ¿ small, the nurse complained that she was not able to advance dilator into sheath.Physician tried as well but it was hard to advance dilator.During visible inspection it became obvious that sheath valve was dislocated into sheath handle and blocked inner lumen.The issue was noticed during preparation of the sheath.A new sheath was used successfully to treat patient without any other issues.Device evaluation details: the product was returned to biosense webster (bwi) for evaluation.Visual inspection and microscopic examination of the returned device were performed following bwi procedures.Visual analysis revealed that the hemostatic valve was dislodged inside of hub component.Microscopic examination of the hemostatic valve surface showed stress marks on the outer diameter.The damage observed could be related to the incorrect insertion of the dilator into the sheath causing the dislodgment of the valve; the stress marks and physical damage observed suggest that excessive force or manipulation was applied; however, this could not be conclusively determined.Additionally, a picture was received for evaluation following biosense webster's procedures.According to the picture provided by the customer, the hemostatic valve was dislodged inside the hub component.This issue could be related to the incorrect insertion of the dilator into the sheath causing the dislodgment of the valve; however, this could not be conclusively determined.The issue reported by the customer was confirmed based on the picture received.A device history record was performed for the finished device batch number, and no internal actions were identified.However, due to the conditions observed in the hemostatic valve, an internal corrective action has been opened to address this issue.The issue reported by the customer was confirmed.The odp (optimal device performance guide) contains the following caution: always insert a dilator straight into the center of the sheath¿s valve to prevent damage to the valve.Do not insert a dilator at an angle, as damage to the sheath valve may occur.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.#(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ SMALL
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (IRVINE)
33 technology drive
irvine CA 92618
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-798
MDR Report Key15577926
MDR Text Key307026737
Report Number2029046-2022-02469
Device Sequence Number1
Product Code DYB
UDI-Device Identifier10846835016253
UDI-Public10846835016253
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K170997
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/10/2023
Device Model NumberD138501
Device Catalogue NumberD138501
Device Lot Number00002054
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received12/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
8.5F SHEATH WITH CURVE VIZ SMC.
-
-