• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP M.R.I. IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP M.R.I. IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 9808560
Device Problems Fracture (1260); Material Separation (1562); Deformation Due to Compressive Stress (2889); Naturally Worn (2988); Migration (4003)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 09/09/2022
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Device pending return.
 
Event Description
It was reported that six years post port placement, the catheter was allegedly broken.It was further reported that distal catheter segment migrated into the heart.Reportedly, the distal catheter segment was removed.The current status of the patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one powerport implantable port attached to a groshong catheter in two segments were returned for evaluation.Functional, gross visual, tactile, and microscopic visual evaluations were performed.The investigation is confirmed for the reported fracture and material separation issues and identified wear and deformation issues, as a partial circumferential break was noted approximately 3.4cm from the distal end of the cath-lock.Complete circumferential breaks were noted on the distal end of the attached catheter and on the proximal end of the distal catheter segment.The surface of the partial circumferential break on the attached catheter was noted to be round and granular.The surfaces of the complete circumferential break on the distal end of the attached catheter and proximal end of the distal catheter segment were noted to be granular and glossy.However, the investigation is inconclusive for the reported migration issue, as the exact circumstances at the time of the reported event cannot be verified, and the reported event could not be reproduced in the lab.The identified break is typical of flexural fatigue, which is due to the repetitive, cyclic kinking of the catheter.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 03/2020).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event Description
It was reported that approximately six years and six months post port placement, the catheter was allegedly broken.It was further reported that distal catheter segment migrated into the heart.Reportedly, the distal catheter segment was removed.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT ISP M.R.I. IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15580582
MDR Text Key301556795
Report Number3006260740-2022-03954
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741027512
UDI-Public(01)00801741027512
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K063377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9808560
Device Catalogue Number9808560
Device Lot NumberREZI0461
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age73 YR
Patient SexFemale
Patient Weight50 KG
-
-