• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. WM CLARIS CPU; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. WM CLARIS CPU; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number H700977
Device Problems Failure to Power Up (1476); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/08/2022
Event Type  malfunction  
Event Description
During device preparation with the patient in the room, a communication issue occurred and the procedure was cancelled.The computer turned off unexpectedly and could not be turned back on.The system was attempted to restarted, the power cable was changed and the power source was changed which did not resolve the issue and the case was cancelled.
 
Manufacturer Narrative
One claris display workstation was received for evaluation.Power was applied to the dws, and the dws passed power-on-self-test (post).The dws loaded the operating system then the workmate claris application successfully.Evaluation of the logs identified that there was soft data corruption.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the investigation and information provided to abbott, the reported event was confirmed, however the root cause was not able to be isolated and the device worked as per designed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WM CLARIS CPU
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15581605
MDR Text Key305605455
Report Number2184149-2022-00227
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05415067009777
UDI-Public05415067009777
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K151911
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH700977
Device Catalogue NumberH700977
Device Lot Number5336605
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-