As reported for this event, during a therapeutic radial ultrasound bronchoscopy (rebus) procedure the blood entered the device.The procedure was completed successfully with the same device without any delay.There is no harm or adverse impact to the patient.Subsequently the device probe was found to be defective with a gap.The device had been inspected prior to use.Per the customer, the gap was created at the device tip because the rotational thread shortens, much more than customer is accustomed to.
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The device was returned and an evaluation was completed for it.During inspection, olympus confirmed the reported event, the probe has a defect and there is also blood in it and there is an ultrasonic medium leak.- it was confirmed that there was blood inside the tip sheath.- it was confirmed that there was a leakage of the ultrasonic medium.- it was confirmed that there was a perforation in the tip sheath.- it was confirmed that the internal blade (flexible shaft) was driven normally.- it was confirmed that the ultrasound image was not drawn normally.Additionally, there is a perforation in the tip sheath and the ultrasound images do not render properly; however, these defects are not considered severe enough to cause a potential adverse event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely an external force caused a perforation in the tip sheath, which allowed blood to get inside.Olympus will continue to monitor field performance for this device.
|