• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES LLC EDWARDS SAPIEN 3 ULTRA; PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES LLC EDWARDS SAPIEN 3 ULTRA; PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9750TFX26
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombosis/Thrombus (4440)
Event Date 03/30/2022
Event Type  Injury  
Event Description
Thv/tvt registry summary reporting for serious injury device thrombosis event associated with the sapien 3 ultra transcatheter heart valve, in the mitral position, for quarter 3 2022 data extract, that occurred more than 1 year post procedure.
 
Manufacturer Narrative
Per the instructions for use (ifu), valve thrombosis is a potential risk associated with the use of the transcatheter heart valve (thv).Valve thrombosis is a rare and well recognized complication associated with the use of bioprosthetic heart valves.Valve thrombosis is the formation of significant blood clots forming on the valve.These clots could significantly impact the functionality of the valve resulting in heart failure or thromboembolism.Anticoagulation therapy must be evaluated patient by patient.The cause for valve thrombosis is thought to result from an interaction between components of blood and the prosthesis or turbulent blood flow in and around the prosthesis.Several factors can cause development of thrombosis, including: inadequate anticoagulant therapy after valve implant, atrial fibrillation, medications, cancerous tumors, systemic diseases (e.G., systemic lupus erythematosus, inflammation and damage to various body tissues, including joints, skin, kidneys, heart, lungs, blood vessels, and brain), and reduced cardiac pumping (low ejection fraction).Thrombosis has an extremely low incidence rate in bioprosthetic heart valves.Thrombosis is a complex process triggered by the interaction between the host and the device and is highly variable among patients.Since the mechanism of thrombotic deposition on bioprosthetic heart valves is not fully understood, the cause cannot always be confirmed.In this case, specific clinical details were not available, and it is not possible to determine potential contributing factors.A complaint history for this type of event is reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventive actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 ULTRA
Type of Device
PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key15592371
MDR Text Key301631114
Report Number2015691-2022-08550
Device Sequence Number1
Product Code NPU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9750TFX26
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age66 YR
Patient SexMale
-
-