• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number 100/543/080CZ
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This mdr was generated under protocol (b)(4)., as a result of warning letter (b)(4).Unique identifier and 510k are unknown.No information has been provided to date.A review of the device history records shows that during manufacturing of this batch 10 new shelf carton labels had to be additionally printed.No product sample was received; therefore, visual and functional testing could not be performed.No visual/functional testing completed on the photos there were 10 shelf cartons within one transit carton without yellow warning sticker "without forceps" plus one shelf carton from the same part number lot number was not visible with yellow sticker "without forceps" for comparison purposes.Based on information provided by customer 10pcs out of 395pcs from lot were found to be without yellow sticker.It was confirm that each shelf carton shall contain yellow sticker "without forceps" therefore manufacturing error was identified.It was the most probable that operators forgot to use yellow labels on those 10 pieces.Therefore we don't expect that more cartons will be affected by missing yellow labels.Improvements which are described in analysis section will prevent to such issues because quantity of used yellow stickers are now fully under control.This customer's complaint was introduced to all employees on monthly town hall meeting.
 
Event Description
It was reported that the round yellow sticker was missing.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TRACHEOSTOMY TUBE
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15604532
MDR Text Key307173355
Report Number3012307300-2022-24014
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number100/543/080CZ
Device Lot Number3637895
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-