• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB ENVISTA INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB ENVISTA INTRAOCULAR LENS Back to Search Results
Model Number MX60E
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Endophthalmitis (1835)
Event Date 08/31/2022
Event Type  Injury  
Event Description
Approximately 8 months post implantation patient developed bilateral precipitates on the anterior of the lens.The surgeon believes the patient has chronic endophthalmitis.
 
Manufacturer Narrative
Device is not available for investigation as it remains implanted in the patient¿s eye.Investigation of this event is in progress.A follow up report will be submitted upon completion of the investigation.
 
Event Description
Additional information from the surgeon confirmed the results of the tap performed were negative.The precipitates disappeared with a steroid with no recurrence.There is currently pigments on the patients lens.
 
Manufacturer Narrative
A review of the device history record (dhr) did not find any non-conformities or anomalies related to the reported event.The trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.Based on all available information the root cause of this event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVISTA INTRAOCULAR LENS
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
BAUSCH + LOMB
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
21 north park place blvd.
clearwater FL 33759
Manufacturer Contact
shayan habibi
21 north park place blvd.
clearwater, FL 33759
7277246600
MDR Report Key15604977
MDR Text Key301723411
Report Number0001313525-2022-00142
Device Sequence Number1
Product Code HQL
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P910056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Model NumberMX60E
Device Catalogue NumberMXUE2550
Device Lot Number3075836
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age78 YR
Patient SexFemale
-
-