Model Number CDVRA500Q |
Device Problems
Inappropriate or Unexpected Reset (2959); Compatibility Problem (2960)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/26/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Further information was requested but not received.
|
|
Event Description
|
During an in-clinic follow-up, the device entered backup vvi (bvvi) due to off-label magnetic resonance imaging (mri).Abbott technical support was contacted, and the device was reprogrammed to resolve the event.The patient was in stable condition.
|
|
Event Description
|
Additional information received indicated mri procedure was followed prior to the procedure and the mri compatible system was programmed to mri mode properly.The bvvi due to mri also resulted in a loss of high voltage therapy.The patient was stable with no adverse consequences.
|
|