Model Number 25052 |
Device Problem
High Readings (2459)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/20/2022 |
Event Type
malfunction
|
Event Description
|
The customer reported the measurements are higher than they expected.No patient impact or consequences were reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative:the device has not been returned to the investigation facility to allow for an analysis to be performed.If the product is received for evaluation or new information is obtained, a follow up report will be submitted.Initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: (b)(6).Device not returned.
|
|
Manufacturer Narrative
|
Additional manufacuring narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed., other, other text: initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: (b)(6).
|
|
Event Description
|
The customer reported the measurements are higher than they expected.No patient impact or consequences were reported.
|
|