Model Number 25052
Device Problem High Readings (2459)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2022
Event Type  malfunction  
Event Description
The customer reported the measurements are higher than they expected.No patient impact or consequences were reported.
 
Manufacturer Narrative
Additional manufacturing narrative:the device has not been returned to the investigation facility to allow for an analysis to be performed.If the product is received for evaluation or new information is obtained, a follow up report will be submitted.Initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: (b)(6).Device not returned.
 
Manufacturer Narrative
Additional manufacuring narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed., other, other text: initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: (b)(6).
 
Event Description
The customer reported the measurements are higher than they expected.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADICAL-7 HANDHELD
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key15610875
MDR Text Key302075632
Report Number3019388613-2022-00224
Device Sequence Number1
Product Code DQA
UDI-Device Identifier00843997008358
UDI-Public00843997008358
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K110028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25052
Device Catalogue Number9500
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received11/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/28/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-