Model Number 212480
Device Problem Residue After Decontamination (2325)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2022
Event Type  malfunction  
Event Description
Oil leaking from angled saw attachment.Case type / application: tka.
 
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Manufacturer Narrative
An event regarding foreign matter involving a mako saw attachment was reported.The event was not confirmed.Method & results product evaluation and results: visual inspection did not confirm the event.Oil was not present.Clinician review: no medical records were received for review with a clinical consultant.Product history review: a product history review is not required, as no manufacturing specific issue has been alleged.At stryker joint replacement all devices/product are manufactured through validated equipment and inspected by trained representatives through stryker¿s quality management system.Complaint history review: a complaint history review and trend detection is not required as this is a pm.Monitoring will be continued under the current quarterly trend review.Conclusions: preventive maintenance is where an action occurs that identifies device deterioration which may compromise function.Under pm conditions no patient was involved and no actual or potential patient harm existed for the alleged event.The alleged failure mode was not confirmed.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
 
Event Description
Oil leaking from angled saw attachment.Case type / application: tka.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7 DEGREE ANGLED SAGITAL SAW
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
juana hiciano
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15624188
MDR Text Key302169978
Report Number3005985723-2022-00132
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00848486032128
UDI-Public00848486032128
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143752
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number212480
Device Catalogue Number212480
Device Lot Number35021116 / 3500381
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received11/14/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-