Model Number 212480 |
Device Problem
Residue After Decontamination (2325)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/04/2022 |
Event Type
malfunction
|
Event Description
|
Oil leaking from angled saw attachment.Case type / application: tka.
|
|
Manufacturer Narrative
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
|
Manufacturer Narrative
|
An event regarding foreign matter involving a mako saw attachment was reported.The event was not confirmed.Method & results product evaluation and results: visual inspection did not confirm the event.Oil was not present.Clinician review: no medical records were received for review with a clinical consultant.Product history review: a product history review is not required, as no manufacturing specific issue has been alleged.At stryker joint replacement all devices/product are manufactured through validated equipment and inspected by trained representatives through stryker¿s quality management system.Complaint history review: a complaint history review and trend detection is not required as this is a pm.Monitoring will be continued under the current quarterly trend review.Conclusions: preventive maintenance is where an action occurs that identifies device deterioration which may compromise function.Under pm conditions no patient was involved and no actual or potential patient harm existed for the alleged event.The alleged failure mode was not confirmed.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
|
Event Description
|
Oil leaking from angled saw attachment.Case type / application: tka.
|
|
Search Alerts/Recalls
|