Catalog Number UNKNOWN |
Device Problems
Structural Problem (2506); Device Tipped Over (2589)
|
Patient Problems
Chest Pain (1776); Headache (1880); Nausea (1970); Internal Organ Perforation (1987); Pain (1994); Perforation of Vessels (2135); Anxiety (2328); Depression (2361); Device Embedded In Tissue or Plaque (3165); Unspecified Eye / Vision Problem (4471)
|
Event Type
Injury
|
Manufacturer Narrative
|
Initial reporter occupation: non-healthcare professional.Investigation: the following allegations have been investigated: organ/vena cava perforation, tilt, anxiety, fear, depression, pain.The reported allegations have been investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported anxiety, fear, depression, pain is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown; however, the alleged filter is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
The following information is alleged: the patient received a celect inferior vena cava (ivc) filter on (b)(6) 2012.Approximately eight years and six months later, the patient underwent a computed tomography (ct) scan which revealed that the filter is angled 28 degrees relative facet ivc with the tip directed anteriorly by about 24 degrees.Also, the tip and multiple struts perforate the wall of the ivc, two struts may enter adjacent liver, one abuts duodenum, and one abuts spine.The patient further alleges anxiety, fear, depression, and pain.Hospital and medical records have been requested, but not yet provided.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.The following fields were updated per additional information received: a2, a4, b5, b6, b7, d6, and h6.Investigation.The following allegations have been investigated: embedded, mood swings, mental challenges, headaches, nausea, chest pains, leg pains, worsened vision.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The additional information regarding embedded does not change the previous investigation results for vena cava/organ perforation.Unknown if the reported mood swings, mental challenges, headaches, nausea, chest pains, leg pains, worsened vision is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown; however, the alleged filter is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Patient allegedly received an implant on (b)(6) 2012 via the right internal jugular vein due to history of pulmonary embolism (pe) during pregnancy.The patient alleges device tilt, vena cava and organ perforation, and embedment.The patient further alleges headaches on and off every month, nausea, mental challenges, chest pains, mood swings, leg pains, and worsened vision.(b)(6) 2020, per a report from computed tomography; ¿vasculature: ivc filter is present.On coronal series, it is angled about 28 degrees relative facet ivc with the tip directed towards the left.On sagittal series, the filter is tilted with its tip directed anteriorly by about 24 degrees.The tip and multiple inferior struts perforate the wall of the ivc.Two of the struts may enter adjacent liver and one abuts duodenum.One of but spine.¿.
|
|
Search Alerts/Recalls
|
|
|