Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problem Pain (1994)
Event Date 09/21/2022
Event Type  malfunction  
Event Description
On (b)(6) 2019 a patient underwent revision surgery to extend their spinal construct from t10-iliac to t8-s2ai as a result of the patients failure to fuse from the previous procedure.On a unknown date in (b)(6) 2022 the patient was experiencing pain and it was discovered that a rod fractured at the l5/s level.On (b)(6) 2022 an additional revision occurred where the fractured rod was removed and replaced.During the procedure loosening to an unknown amount of screws was also observed.The loosened screws were replaced with upsized screws and the construct was extend to t7-il.Post revision it was reported fusion had again not been achieved and was suspect to be a contributing factor of the rod fractures.
 
Manufacturer Narrative
No product was returned and no radiographs were provided confirming the alleged complaint.It is unknown what the patients post activity levels were or whether they experienced a fall or other accident.Review of the reported event identified the patient has a history of spinal procedures and pseudarthrosis.It is unknown if the patient is a smoker.Review of all provided information indicates the root cause of the loosened screws to be the result of an extended period (3 plus years) of non fusion resulting in rod fracture excessive loading and subsequent screw loosening.No additional investigation required.Labeling review: "contraindications - contraindications include, but are not limited to: patients with inadequate bone stock or quality." "potential adverse events and complications - as with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries." "potential risks identified with the use of this system, which may require additional surgery, include" bending, fracture or loosening of implant component(s), loss of fixation, nonunion or delayed union." "warnings, cautions and precautions - the safety and effectiveness of this device has not been established for use as part of a growing rod construct.This device is only intended to be used when definitive fusion is being performed at all instrumented levels." "these devices can break when subjected to the increased load associated with delayed union or nonunion.Internal fixation appliances are load-sharing devices that hold bony structures in alignment until healing occurs.If healing is delayed, or does not occur, the implant may eventually loosen, bend, or break.Loads on the device produced by load bearing and by the patient¿s activity level will dictate the longevity of the implant." "patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." "post-operative warnings - during the postoperative phase it is of particular importance that the physician keeps the patient well informed of all procedures and treatments.Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk blvd
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk blvd
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk blvd
san diego, CA 92121
MDR Report Key15641374
MDR Text Key306647398
Report Number2031966-2022-00214
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K143684
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberNI
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
-
-