MAUDE Adverse Event Report: INSULET CORPORATION OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSULET CORPORATION OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN Back to Search Results
Model Number PT-000010
Device Problem Failure to Reset (1532)
Patient Problem Insufficient Information (4580)
Event Date 10/15/2022
Event Type  Injury  
Manufacturer Narrative
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to determine if any product condition could have contributed to the reported hospitalization.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including communicating with the pdm, deployment, delivering fluid, occlusion detection, and freedom from hazard alarms.
 
Event Description
It was reported that the patient visited the emergency room (er) due to not being able to control the blood glucose (bg0 levels as the personal diabetes manager (pdm) stopped working caused by water damaged and could not be turned on to operate the pod.At the hospital, the patient was kept overnight to be observed until a new pdm is given and insulin was administered manually.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
michael spears
100 nagog park
acton, MA 01720
9786007000
MDR Report Key15649497
MDR Text Key302167413
Report Number3004464228-2022-19726
Device Sequence Number1
Product Code LZG
UDI-Device Identifier10385082000139
UDI-Public(01)10385082000139(11)210225(10)7560
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 10/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberPT-000010
Device Catalogue NumberINT1-D001-MG
Device Lot Number7560
Was Device Available for Evaluation? No
Date Manufacturer Received10/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/25/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age11 YR
Patient SexMale
-
-