SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA BAGS; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
|
Back to Search Results |
|
Catalog Number 670001 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/24/2018 |
Event Type
malfunction
|
Event Description
|
It was reported when the customer attempted to perform air ventilation at a pre-use check on the product, the doctor noticed the bag was torn.During the air leak check, the product worked normally.It is not certain why the product passed the air leak check in spite that the bag was torn.No patient injury reported.
|
|
Manufacturer Narrative
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.The sample was received in used conditions, without its original packaging.The sample was visually inspected at a distance of 12" under normal lighting to received units, in order to detect any damage on the unit.The breathing bag is torn.Production personnel did not detect and scraped the damaged breathing bag.Actions taken to mitigate the issue: production personnel was notified by the quality engineer on the failure mode reported by the customer.Udi is unknown.No product information has been provided to date.
|
|