Model Number D138501
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/26/2022
Event Type  malfunction  
Manufacturer Narrative
The bwi product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that an unknown patient underwent an atrial fibrillation (afib) ablation procedure with a carto vizigo¿ 8.5f bi-directional guiding sheath ¿ small.The hemostatic valve was leaking.The hemostatic valve on the vizigo sheath was leaking back blood.The vizigo sheath was replaced, and the issue resolved.The procedure continued.It was reported that the hemostasis valve did not dislodge inside the hub but was leaking.The brim cap/hub did not detach from the sheath.It was not stated by physician if air entered the patient¿s body.No patient consequences.The approximate volume of blood that was lost was unknown but enough for physician to change sheath.Hemostatic valve leak is mdr-reportable.
 
Manufacturer Narrative
On 5-nov-2022, the product investigation was completed.It was reported that an unknown patient underwent an atrial fibrillation (afib) ablation procedure with a carto vizigo¿ 8.5f bi-directional guiding sheath ¿ small.The hemostatic valve was leaking.Device evaluation details: visual analysis revealed that the hemostatic valve was dislodged inside the hub component.Microscopic examination of the hemostatic valve surface showed stress marks on the outer diameter.The stress marks suggest that excessive force or manipulation was applied due to an extreme off axis angle of insertion.Valve dislodgement occurs when extreme off axis angles are performed during insertion with the dilator, outside of what is recommended in the odp (optimal device performance guide).A device history record evaluation was performed for the finished device number 00002036 and no internal action was found during the review.The issue reported by the customer was confirmed.The odp contains the following caution: always insert a dilator straight into the center of the sheath¿s valve to prevent damage to the valve.Do not insert a dilator at an angle, as damage to the sheath valve may occur.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH ¿ SMALL
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (IRVINE)
33 technology drive
irvine CA 92618
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key15656884
MDR Text Key307203747
Report Number2029046-2022-02613
Device Sequence Number1
Product Code DYB
UDI-Device Identifier10846835016253
UDI-Public10846835016253
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170997
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/17/2023
Device Model NumberD138501
Device Catalogue NumberD138501
Device Lot Number00002036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received11/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-