Model Number SPD2-US-060-320 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 08/01/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Related to 3600510000-2022-8009 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
A physician was using a spider fx embolic protection device for treatment of the popliteal artery.It was reported that the spider f ilter was advanced. the non-mdt support catheter was then exchanged for a 2.4mm non-mdt atherectomy catheter.One pass was made in the proximal part of the occluded sfa. after the catheter was pulled back up, the filter migrated up as well. attempts to advance the filter resulted in separation of the proximal part of the wire from the distal part.The non-mdt catheter was removed, and the non-mdt support catheter was readvanced into the popliteal artery.A non-mdt wire was then advanced into the popliteal after next to the spider filter.The support catheter was removed, and the filter was then retrieved.The support catheter was readvanced over the non-mdt wire and the wire was exchanged for a new filter.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|