OGDEN MANUFACTURING PLANT OPTIFLUX 160NRE DIALYZER FINISHED ASSY.; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Model Number 0500316E |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/19/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
|
Event Description
|
A user facility reported that a blood leak occurred immediately after the start of the patient¿s hemodialysis (hd) treatment.Upon follow-up, the clinic manager (cm) stated the machine, a fresenius 2008k2 machine, alarmed appropriately with a "minor" blood leak alarm.Blood leak test strips were used and tested positive for the presence of blood.The cm also stated that fresenius bloodlines were used for treatment and confirmed that the leak was internal and no external blood leak was observed.The cm stated that there were no defects or damage seen on the dialyzer.Per cm the patient¿s blood was not returned and stated the estimated blood loss was 150 ml.Immediately following the event, the patient was re-setup with new supplies on a different machine where the patient was able to complete treatment.The cm confirmed that there was no patient injury, adverse events, or medical intervention required as a result of the reported event.The dialyzer was available to be returned for manufacturer evaluation.
|
|
Event Description
|
A user facility reported that a blood leak occurred immediately after the start of the patient¿s hemodialysis (hd) treatment.Upon follow-up, the clinic manager (cm) stated the machine, a fresenius 2008k2 machine, alarmed appropriately with a "minor" blood leak alarm.Blood leak test strips were used and tested positive for the presence of blood.The cm also stated that fresenius bloodlines were used for treatment and confirmed that the leak was internal and no external blood leak was observed.The cm stated that there were no defects or damage seen on the dialyzer.Per cm the patient¿s blood was not returned and stated the estimated blood loss was 150 ml.Immediately following the event, the patient was re-setup with new supplies on a different machine where the patient was able to complete treatment.The cm confirmed that there was no patient injury, adverse events, or medical intervention required as a result of the reported event.The dialyzer was available to be returned for manufacturer evaluation.
|
|
Manufacturer Narrative
|
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned to date for manufacturer evaluation.A lot history review was performed.There was one other complaint reported against the lot.The complaint is addressing a dialyzer that had an internal blood leak which was reportedly seen within the header and the dialysate effluent (no sample).A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was one approved temporary deviation notice (dn) reported on the lot which was unrelated to the complaint event.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.The reported lot number passed pyrogen testing, was within sterilization dosage parameters and passed all release criteria.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas 2018-0171 (vision systems) and 2018-0161 (blood leak reduction) are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
|
|