Model Number DTMA1QQ
Device Problems Failure to Deliver Shock/Stimulation (1133); Device Sensing Problem (2917)
Patient Problems Dyspnea (1816); Inflammation (1932); Nausea (1970); Tachycardia (2095); Diaphoresis (2452); Pallor (2468)
Event Date 12/28/2021
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient presented to emergency department complaining of chest pain, shortness of breath, nausea, feeling warm and sweating.The patient collapsed into ventricular tachycardia (vt) and was externally shocked three times.It was noted that the patient was hypotensive, pale in color with heart rhythm in atrial fibrillation (af).The patient was found coronavirus disease (covid-19).The patient was hospitalized due to vt.The patient went into ventricular fibrillation (vf) for which the crt-d appropriately detected and terminated with one administered high voltage therapy.The patient was administered antiarrhythmic, magnesium sulfate, antiemetic, and norepinephrine medication intravenously.The crt-d was re-programmed to increase specificity of supra ventricular tachycardia (svt) decimators.The crt-d remains in use.No further patient complications have been reported as a result of this event.The patient is a participant in a clinical study.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARIA MRI QUAD CRT-D SURESCAN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key15674652
MDR Text Key302431281
Report Number9614453-2022-03319
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00763000059514
UDI-Public00763000059514
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2022
Device Model NumberDTMA1QQ
Device Catalogue NumberDTMA1QQ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/27/2022
Date Device Manufactured09/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
429888 LEAD, 6947M55 LEAD, 5076-45 LEAD
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age73 YR
Patient SexFemale
Patient Weight87 KG
-
-