• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID W/ E/F PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID W/ E/F PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-55
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Appropriate Term/Code Not Available (3191)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2022
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during service, the cardiosave intra-aortic balloon pump (iabp) was determined that the drive manifold test was out of tolerance.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and was able to confirm the reported issue, drive manifold leak detected.The fse replaced the drive manifold to resolve the issue.Inspection carried out.All measured values ¿¿are in the manufacturer's specifications.The fse handed over equipment to customer and was put back into operation.
 
Event Description
It was reported that during functional test/maintenance, the cardiosave intra-aortic balloon pump (iabp) was discovered/determined that the drive manifold test was out of tolerance.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID W/ E/F PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15674927
MDR Text Key306929051
Report Number2249723-2022-02764
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108414
UDI-Public10607567108414
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-55
Device Catalogue Number0998-00-0800-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-