Lot Number ASKU
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2022
Event Type  malfunction  
Event Description
It was reported that a leak was observed while disinfecting a u9000 ultrafilter.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Initial reporter address: (b)(6).Initial reporter phone no.(b)(6).The device was evaluated onsite by the baxter technician and confirmed that the elbow connector of evacuation valve (evva) was disconnected and the ultrafilter was leaking.To resolve the issue, the qualified technician put the connection back in position firmly and replaced the ultrafilter.Based on past investigations, the most likely failure is a crack in the housing nearby the welding zone.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
U9000
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
holger - crafoord - strasse 26
hechingen D-723 79
GM   D-72379
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15675285
MDR Text Key305219236
Report Number9611369-2022-00203
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/01/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
AK 96 MACHINE
-
-