Catalog Number 102297 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/03/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during treatment with a phoenix/ innova machine, a fluid leak was observed.No alarm was triggered.Treatment was discontinued and the patient was switched to another machine.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The device was evaluated on site by a qualified technician.Upon visual inspection a fluid leakage from the isolation valve (ev1) and hemodialyzer outlet solenoid valve (ev2) was observed.The reported condition was verified.The valves were replaced, and no further issues were noted.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
|
|