• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number 100/563/090
Device Problem Incomplete or Missing Packaging (2312)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/30/2018
Event Type  malfunction  
Manufacturer Narrative
This mdr was generated under protocol (b)(4) as a result of warning letter cms# (b)(4).A manufacturing device history record (dhr) review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.A product sample was received for evaluation.Visual and functional testing were performed.Physical condition of the device showed samples were accompanied with a certificate of safe handling and were not in the original packaging.Customer?s reported issue was duplicated.It was established that the customer received the wrong introducer piece.The root cause of the reported issue was determined.Incorrect alignment of the tube with the flange as well as lack of detection by the production personnel.Actions taken to mitigate the reported issue: review of the operators instructions for kitting products and the assembly packaging instructions was performed.The assembly packaging instructions does state that the size of the introducer piece is required to be checked prior to build, there are no clear instructions in the kitting procedure to emphasis this check as a critical part of the process.Udi information is unknown.Protocol # premarket (510k) number is unknown.
 
Event Description
It was reported that the purple introducer is too thin and too short for the 9.0 cannula.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TRACHEOSTOMY TUBE
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15685449
MDR Text Key306919734
Report Number3012307300-2022-26387
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number100/563/090
Device Lot Number3566236
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2018
Was the Report Sent to FDA? No
Date Manufacturer Received09/26/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-