Catalog Number UNKNOWN- DIALYZER |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2022 |
Event Type
malfunction
|
Event Description
|
A facebook social media user reported that a dialyzer blood leak occurred during a patient's hemodialysis (hd) treatment.It is unknown what type of dialyzer was being used.It was reported that the blood leak was caught almost immediately, and as a result the machine did not alarm.No further details were provided.
|
|
Manufacturer Narrative
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
|
Event Description
|
A facebook social media user reported that a dialyzer blood leak occurred during a patient's hemodialysis (hd) treatment.It is unknown what type of dialyzer was being used.It was reported that the blood leak was caught almost immediately, and as a result the machine did not alarm.No further details were provided.
|
|
Manufacturer Narrative
|
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.In addition, the catalog and lot numbers were not provided.A shipping history search for dialyzers delivered to the account could not be performed because an account number was not identified.As such, device history and manufacturing reviews could not be conducted.Due to the limited information available, an investigation could not be performed.A definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
|
|
Search Alerts/Recalls
|