• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US VAPR VUE WIRELESS FOOTSWITCH; FOOT-SWITCH, ELECTRICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VAPR VUE WIRELESS FOOTSWITCH; FOOT-SWITCH, ELECTRICAL Back to Search Results
Model Number 227214
Device Problems Break (1069); Use of Device Problem (1670); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2022
Event Type  malfunction  
Event Description
It was reported by a sales rep that during a shoulder repair procedure on (b)(6) 2022, it was observed that the back of the vapr vue wireless footswitch device was coming off.According to the report, the bottom was attached by one screw but the rest were off.During in-house engineering evaluation, it was determined that the device was broken/fractured.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The expiration date is currently unavailable.The device manufacture date is currently unavailable.Investigation summary: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: broken (2+ pieces) : broken/fractured.Per service reports, this complaint can be confirmed.The repair of the device was however declined, and it is being placed into long term hold.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR VUE WIRELESS FOOTSWITCH
Type of Device
FOOT-SWITCH, ELECTRICAL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15695047
MDR Text Key306884880
Report Number1221934-2022-03341
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10886705009657
UDI-Public10886705009657
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number227214
Device Catalogue Number227214
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/14/2022
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-