C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP M.R.I. IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 1806060J |
Device Problems
Fracture (1260); Air/Gas in Device (4062)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/03/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a port placement procedure, a crack allegedly formed on the hub of the introducer needle.It was further reported that air was allegedly aspirated during aspiration.There was no reported patient injury.
|
|
Manufacturer Narrative
|
The catalog number has not been cleared in the us but is similar to the powerport isp m.R.I.Implantable port, chronoflex single-lumen, 6f products that are cleared in the us.The pro code and 510 k number for the powerport isp m.R.I.Implantable port, chronoflex single-lumen, 6f products are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 07/2023).
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the powerport isp m.R.I.Implantable port, chronoflex single-lumen, 6f products that are cleared in the us.The pro code and 510 k number for the powerport isp m.R.I.Implantable port, chronoflex single-lumen, 6f products are identified in d2 and g4.H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one introducer needle was returned for evaluation.Gross, microscopic visual and functional testing were performed.The investigation is confirmed for the reported air aspiration and fracture issues as cracks were noted throughout the needle hub.During functional testing, upon infusion, leak was observed from the introducer needle hub and aspiration was attempted but was unsuccessful.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 07/2023).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a port placement procedure, a crack allegedly formed on the hub of the introducer needle.It was further reported that air was allegedly aspirated during aspiration.There was no reported patient injury.
|
|