• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, CROSSFIRE 2 CONSOLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, CROSSFIRE 2 CONSOLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0475100000
Device Problems Thermal Decomposition of Device (1071); Activation Failure (3270)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2022
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was a thermal event.
 
Manufacturer Narrative
Alleged failure: just got this unit back from repair, tried to use it and it smelled like something was burning upon first use.Salesforce case number (b)(4).The failure(s) identified in the investigation is consistent with the complaint record.The root cause is the rf board.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Event Description
It was reported that there was a thermal event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, CROSSFIRE 2 CONSOLE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15708742
MDR Text Key302797573
Report Number0002936485-2022-00624
Device Sequence Number1
Product Code GEI
UDI-Device Identifier07613327058109
UDI-Public07613327058109
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071859
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0475100000
Device Catalogue Number0475100000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received10/05/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-