Catalog Number UNKNOWN
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Muscular Rigidity (1968)
Event Date 07/09/2020
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that on the literature review "plate fixation versus arthroscopic-assisted plate fixation for isolated medium-sized fractures of the greater tuberosity: a retrospective study", 1 patient had postoperative shoulder stiffness after an arthroscopic-assisted plate fixation procedure using two (2) twinfix ti anchors.The patients was manipulated to release adhesion under anesthesia.The patient obtained satisfactory shoulder function after manipulation, based on the self-evaluation.No further information is available.
 
Manufacturer Narrative
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product identification information was provided and thus a manufacturing record review, complaint history review, risk management review, and an instructions for use/device labeling review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A clinical review states without clinically relevant patient-specific supporting documentation, a thorough medical investigation cannot be performed.The root cause and/or patient outcome beyond that which was documented in the article cannot be confirmed nor concluded.In addition, the physician referenced in the abstract provided an analysis of all of the attached images.Therefore, no further interpretation of the attached images are required.No further medical assessment is warranted at this time.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN METAL SUTURE ANCHOR
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15726293
MDR Text Key303030299
Report Number1219602-2022-01682
Device Sequence Number1
Product Code MBI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K093228
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Study,Literature,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-