• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751761
Device Problem Insufficient Flow or Under Infusion (2182)
Patient Problems Eye Burn (2523); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/06/2022
Event Type  Injury  
Event Description
The nurse reported that corneal wound appeared to be burned at the end of the cataract with iol (intraocular lens) implantation and would not seal.The surgeon had to place 3 (10-0) nylon sutures to ensure closure, and the surgeon commented on a decrease in flow.The details about patient impact was not reported.Additional information had been requested.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Additional information provided in sections h.6.And h.10.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.There was no equipment returned on this investigation.Based on the information obtained, the root cause of the reported event is inconclusive.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15736143
MDR Text Key303089037
Report Number2028159-2022-01567
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657517619
UDI-Public00380657517619
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065751761
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
3 (10-0) NYLON SUTURES; CENTURION VISION SYSTEM
Patient Outcome(s) Other; Required Intervention;
-
-