B. BRAUN SURGICAL, S.A. MONOMEND MAX 2/0 27" (70CM)HR26S VPL RCP; SYNTHETIC ABSORBABLE MONOFILAR SUTURE Back to Search Results
Model Number 100523598
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  malfunction  
Event Description
It was reported an issue with monomend max suture.The client (veterinarian) reported that had a pack 2 months ago and one again yesterday with needles coming off during surgery.No more information has been provided.
 
Manufacturer Narrative
Reported device marketed in the u.S.Only for veterinary use, however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.K031216.Analysis and results: there are no previous complaints of this code batch of which we manufactured and distributed (b)(4) units of this code batch, there are no units in our stock.Without closed and/or defective samples a proper analysis cannot be performed.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done and the case is not confirmed due to lack of evidence.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyse it.Please note that when no samples are received our analysis is very limited.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOMEND MAX 2/0 27" (70CM)HR26S VPL RCP
Type of Device
SYNTHETIC ABSORBABLE MONOFILAR SUTURE
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
MDR Report Key15741902
MDR Text Key307675885
Report Number3003639970-2022-00489
Device Sequence Number1
Product Code NEW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number100523598
Device Catalogue Number100523598
Device Lot Number119515
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-