There were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 2119782.Medical device expiration date: 2023-04-30.Device manufacture date: 2022-05-05.Medical device lot #: 2087723.Medical device expiration date: 2023-03-31.Device manufacture date: 2022-04-01.Investigation summary: bd had not received samples, but (b)(6) photos were provided for investigation.The photos were reviewed and the indicated failure mode for tube breakage was observed.Additionally, (b)(6) retention samples from each batch were evaluated by functional testing and the issue of tube breakage was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode tube breakage.Bd was not able to identify a root cause for the indicated failure mode.
|
It was reported when using the bd vacutainer® sst blood collection tube there was glass breakage during use and sample leakage.The following information was provided by the initial reporter.The customer stated: "the outpatient blood vessel is transported through the airway to the laboratory, then the tube is centrifugally blasted and broken after pretreatment.There was blood exposure in this incident, but no contact between the medical staff and blood.The medical staff wore protective equipment.".
|