• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MAKO ROBOTIC ARM 3.1; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MAKO ROBOTIC ARM 3.1; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 219999
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2022
Event Type  malfunction  
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event Description
Mps says the robot will get into haptics and start to oscillate so much it kicks itself out of haptics.
 
Manufacturer Narrative
Reported event an event regarding unintended movement involving a mako robotic arm was reported.The event was not confirmed.Method & results: product evaluation and results: the field service engineer reported: problem reproduced? no.Trouble shooting notes: robot ¿ mps says the robot will get into haptics and start to oscillate so much it kicks itself out of haptics.Work performed: adjusted j5, j6, j4.Work order disposition: system investigation completed successfully as per service manual.All system checks and tests passed, system is ready for use.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there are no other complaints with a similar failure mode for this lot.Conclusions: the alleged failure mode was not confirmed.Successfully completed all checks, verifications, and calibrations.System is ready for clinical use.
 
Event Description
Mps says the robot will get into haptics and start to oscillate so much it kicks itself out of haptics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAKO ROBOTIC ARM 3.1
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
perla zima
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key15776243
MDR Text Key304888668
Report Number3005985723-2022-00147
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142530
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number219999
Device Catalogue Number219999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-