• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH LOCKING SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L16MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH LOCKING SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L16MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 657316
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2022
Event Type  malfunction  
Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event Description
It was reported that there was metal debris after removing the plate.
 
Manufacturer Narrative
Please note the correction made to the h6 component code: the reported event was confirmed.The device inspection revealed the following: heavy damage can be observed on the screw¿s tip, most probably due to explantation, but also to a possible excessive angulation of the screw, out of the advised range (-15° to 15°) during insertion.The damage observed led to the for-mation of fragments.The slight decolorization and the wear observed on the bottom of the screw head give indi-cation that the screw was overtightened on the plate.Based on investigation, the root cause was attributed to a user related issue.Evidence of screw overtightening and unadvised screw angulation were observed, which should be avoided as it could damage the implant.As a reminder, the instructions for use clearly state that: "¿ screws must not be over-tightened during insertion.Excessive overtightening will compromise the integrity of the screw head, resulting in possible screw breakage." a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Event Description
It was reported that there was metal debris after removing the plate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L16MM
Type of Device
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15796156
MDR Text Key307325049
Report Number0008031020-2022-00607
Device Sequence Number1
Product Code HWC
UDI-Device Identifier07613327086874
UDI-Public07613327086874
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K180500
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number657316
Device Catalogue Number657316
Device Lot NumberH33194
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-