Model Number 1258T/86
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2022
Event Type  malfunction  
Event Description
It was reported that the patient presented for implant procedure.During the procedure, the left ventricular lead failed to advance over the guidewire.The lead was not used and replaced.The patient was recovering post procedure.
 
Manufacturer Narrative
The reported event of the guidewire failure to advance was confirmed.The final analysis found that as received, a complete lead without the guidewire was returned in one piece for analysis.The guidewire insertion test was performed and found the test guidewire was restricted distal to the connector boot.Destructive analysis of the lead found the inner coil was clogged with blood/body fluid at the guidewire restriction location.The cause of the reported event was isolated to the inner coil being clogged with blood/body fluid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX MICRO LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15798519
MDR Text Key303732962
Report Number2017865-2022-45428
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734502610
UDI-Public05414734502610
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1258T/86
Device Catalogue Number1258T-86
Device Lot NumberA000119959
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial