• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US VAPR TRIPOLAR90 SUCTION ELECTRODE; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VAPR TRIPOLAR90 SUCTION ELECTRODE; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE Back to Search Results
Model Number 225028
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2022
Event Type  malfunction  
Event Description
It was reported by the affiliate in japan that during an arthroscopic rotator cuff repair procedure on (b)(6) 2022, it was observed that the suction function on the vapr tripolar90 suction electrode was not working.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Reporter is a j&j employee.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available or the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary : the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number (u2202041), and no non-conformance was identified.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR TRIPOLAR90 SUCTION ELECTRODE
Type of Device
ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15799269
MDR Text Key307675152
Report Number1221934-2022-03570
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10886705023103
UDI-Public10886705023103
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K143475
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number225028
Device Catalogue Number225028
Device Lot NumberU2202041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-