Model Number A219 |
Device Problems
No Audible Alarm (1019); Premature Discharge of Battery (1057); Interrogation Problem (4017)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/02/2022 |
Event Type
Injury
|
Event Description
|
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) had difficulty being interrogated during an s-icd replacement procedure for premature battery depletion.Tones were not being emitted with magnet application.The s-icd was successfully replaced with a new device, and the electrode remains in service.No additional adverse patient effects were reported.
|
|
Event Description
|
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) had difficulty being interrogated during an s-icd replacement procedure for premature battery depletion.Tones were not being emitted with magnet application.The s-icd was successfully replaced with a new device, and the electrode remains in service.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, this s-icd was thoroughly inspected and analyzed.A high current drain was detected.Detailed analysis confirmed the identified high current drain was a result of a compromised capacitor.This resulted in the observed premature battery depletion.It was determined that the capacitor was compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of emblem devices that have a potential of exhibiting this behavior.This particular device is included in the emblem s-icd accelerated depletion advisory population.
|
|
Search Alerts/Recalls
|