Model Number A219
Device Problems No Audible Alarm (1019); Premature Discharge of Battery (1057); Interrogation Problem (4017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2022
Event Type  Injury  
Event Description
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) had difficulty being interrogated during an s-icd replacement procedure for premature battery depletion.Tones were not being emitted with magnet application.The s-icd was successfully replaced with a new device, and the electrode remains in service.No additional adverse patient effects were reported.
 
Event Description
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) had difficulty being interrogated during an s-icd replacement procedure for premature battery depletion.Tones were not being emitted with magnet application.The s-icd was successfully replaced with a new device, and the electrode remains in service.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this s-icd was thoroughly inspected and analyzed.A high current drain was detected.Detailed analysis confirmed the identified high current drain was a result of a compromised capacitor.This resulted in the observed premature battery depletion.It was determined that the capacitor was compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of emblem devices that have a potential of exhibiting this behavior.This particular device is included in the emblem s-icd accelerated depletion advisory population.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM MRI S-ICD
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15802039
MDR Text Key303699888
Report Number2124215-2022-47322
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526584404
UDI-Public00802526584404
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110042/S058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup
Report Date 11/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/29/2018
Device Model NumberA219
Device Catalogue NumberA219
Device Lot Number202817
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/02/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0936-2021
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age72 YR
Patient SexMale
-
-