It was reported that on (b)(6) 2022, a 25mm navitor valve was selected for implantation using a small flexnav delivery system.During the procedure, while deploying the valve, the valve failed to separate from the delivery system due to a retaining tab remaining attached.It was attempted to retract the valve, but that was unsuccessful.An attempt to rotate the delivery system slightly was also unsuccessful.The delivery system was unable to be retracted from patient anatomy, and a sternotomy was performed to explant the delivery system and valve.A new 25mm epic valve was implanted successfully.The patient is reported to be stable.The patient remained hemodynamically stable throughout the procedure.No additional information was provided.
|
An event of the valve failing to separate from the flex nav delivery system due to a retaining tab remaining attached was reported.The device was returned to abbott, and the investigation confirmed that the retainer tabs of the valve were bent.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications, including passing inspections to confirm there was no damage or bends to the valve stent.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with respect to the labeling design or manufacturing of the device.
|