MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL EXPRESSEW III AUTOCAPTURE + SUTURE PASSER; SUTURE/NEEDLE PASSER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL EXPRESSEW III AUTOCAPTURE + SUTURE PASSER; SUTURE/NEEDLE PASSER, REUSABLE Back to Search Results
Model Number 288233
Device Problems Mechanical Problem (1384); Mechanical Jam (2983); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/27/2022
Event Type  malfunction  
Event Description
It was reported that the expressew iii autocapture + suture passer device had an unspecified malfunction.During in-house engineering evaluation, the device was tested for its functionality where an eiii needle test was loaded into the device and was tested on a sample rubber strip.According to the investigation, when the trigger was actuated to deploy the needle, there was a slight resistance, and the deployment was rough causing stuck issue; and that to restore it to the original position, the needle trigger must be pushed back manually.There was no procedure nor patient involvement reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Reporter is a j&j sales representative.The device manufacture date is currently unavailable.The product was returned to (b)(4) for evaluation.(b)(4) then conducted visual inspection and functional test of device received visual observations revealed signs of use but in expected condition.To test its functionality, an eiii needle test was loaded into the device and was tested on a sample rubber strip.When the trigger was actuated to deploy the needle, there was a slight resistance, and the deployment was rough causing stuck issue.To restore it to the original position, the needle trigger must be pushed back manually.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.The possible root cause for deployment issue can be attributed an improper maintenance would lead to bio-debris build up inside the expressew shaft causing wear of internal components leading to rough deployment issues.As per ifu, it is important to inspect the device prior to use to ensure proper mechanical function.Also, it is necessary to follow the instructions to cleaning and sterilization process and between uses, lubricate moving parts with the water-solubel lubricant in accordance with the manufacturer¿s instructions.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESSEW III AUTOCAPTURE + SUTURE PASSER
Type of Device
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15815358
MDR Text Key307188544
Report Number1221934-2022-03614
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10886705027101
UDI-Public10886705027101
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number288233
Device Catalogue Number288233
Device Lot Number66267-220607-07
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/06/2022
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-