Catalog Number 829155 |
Device Problems
Material Frayed (1262); Product Quality Problem (1506); Manufacturing, Packaging or Shipping Problem (2975); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/17/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Initial trend analysis for lot 56711a was conducted, no malfunctions were found.This is the only complaint for lot 56711a.Further investigation will be conducted to determine the root cause of complaint.
|
|
Event Description
|
End user reports feelings burrs on the needle as well as being able to easily remove the plunger from the barrel.
|
|
Manufacturer Narrative
|
No device was returned for testing.Retained lot number 56711a investigated and no malfunctions were found during inspection.No abnormalities were detected.
|
|
Event Description
|
End user reports feelings burrs on the needle as well as being able to easily remove the plunger from the barrel.
|
|