C.R. BARD, INC. (BASD) -3006260740 M.R.I. LOW-PROFILE IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 7F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 0603870CE |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/21/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The catalog number identified has not been cleared in the us, but is similar to the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products that are cleared in the us.The pro code and 510k number for the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products is identified in common device name and pma/510k.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Udi number (expiry date: 09/2026).
|
|
Event Description
|
It was reported that prior to port placement procedure, catheter allegedly fractured.There was no patient contact.
|
|
Event Description
|
It was reported that prior to port placement procedure, catheter allegedly fractured.There was no patient contact.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products that are cleared in the us.The pro code and 510k number for the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products is identified in d2 and g4.H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one groshong catheter loaded with one cath-lock and a catheter stylet were returned for evaluation.Functional and gross visual evaluations were performed.The investigation is unconfirmed for the reported catheter fracture issue, as the port body was patent to infusion and aspiration without issue and no fracture was identified.Also, the split mentioned is the valve which will be present in the groshong port single lumen as per specifications.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 09/2026), g3.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Search Alerts/Recalls
|
|
|