Model Number PM2272 |
Device Problems
Intermittent Capture (1080); Migration or Expulsion of Device (1395); Under-Sensing (1661); Device Sensing Problem (2917)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/01/2022 |
Event Type
Injury
|
Event Description
|
Related manufacturer reference number: 2017865-2022-46038, 2017865-2022-46039.It was reported that intermittent capture, decreased r-wave sensing and varying pacing impedance were observed on the right ventricular (rv) channel.Additionally, decreased p-wave sensing, noise and oversensing were observed on the right atrial (ra) channel.The device was reprogrammed as an interim solution.An x-ray was performed and the device was found to have migrated from the original implanted location.A device revision procedure was performed, the device and rv lead were explanted and replaced to resolve the event.The patient was in stable condition.
|
|
Manufacturer Narrative
|
The event of failure to capture, sensing issue, and device migration was not confirmed.The device was received from the field with battery voltage above elective replacement level.Electrical and mechanical analysis performed indicated normal functionality.Additionally, visual inspection found no anomaly on the header related to the field concern.Further atrial, and ventricular capture test under nominal and field conditions indicated normal results.Longevity assessment was performed, and the device was in normal range of operation with appropriate remaining longevity.
|
|
Search Alerts/Recalls
|
|